Manual for LoTHR Electronics Station
The Electric Load XY-FZ35
Settings
Protection Settings
OVP | overvoltage Protection (default 25.2V) |
OCP | overflow Protection (default 4.10A/5.10A) |
Second, the electronic load on the initial state of power:
The electronic load automatically saves the switching state before shutting down and acts as the initial state of the next power-on;
Third, discharge capacity and discharge time statistics:
Discharge capacity Statistics: the load current is not zero start statistics, until the next load current is zero, it is believed that the completion of a discharge process, statistical end; Discharge time Statistics: from the current is not zero start statistics, to the next load current is zero, think that the completion of a test process, statistical end;
Four, set the maximum capacity (OAH) and set the maximum discharge time (OHP):
Set maximum Capacity (OAH): When the OAH function is started, when the load discharge capacity is higher than the maximum capacity set, the electronic load stops automatically and flashes to display "OAH", and when the OAH alarm is lifted, the capacity statistics are automatically emptied; Set the maximum discharge time (OHP): When the OHP function is started, when the load running time is greater than the maximum discharge time set, the electronic load stops automatically and flashes to display "OHP", and when OHP alarm is lifted, the time statistics are automatically emptied;
Note: In the absence of the OAP and OHP function, the electronic load will record the discharge capacity and discharge time, turn on the OAH and OHP function, to the set value after the electronic load stopped working; When the OHP function is turned on, the running time of the electronic load is the countdown mode;
OAP and OHP functions, can be a good implementation of unattended power aging testing.
Five, Data group Features:
Divided into “DAT0”, “DAT1”
DAT0: Flashing shows only the capacity value and running time of the previous stage, and does not add up to the next stage;
DAT1: Flashing shows the capacity value and running time of the previous stage, and automatically adds up to the next stage;
Serial Interface
Connect your PC via USB port above FZ35. Use serial console with 9600 8 N 1
Serial Commands | |
---|---|
start | Start upload |
stop | Stop upload |
on | Turn on Load features |
off | Turn off Load features |
x.xxA | Set load current |
LVP:xx.x | Set down Undervoltage |
OVP:xx.x | Set over pressure |
OCP:x.xx | Set over flow |
OPP:xx.xx | Set over power |
OAH:x.xxx | Set maximum capacity |
OHP:xx:xx | Set maximum discharge time |
read | Read product parameter settings |
Data upload format
No alarm status:
Load supply voltage, load current, capacity value, discharge time;
11.90V,0.11A,0.004Ah,00:02 à Load voltage 11.90V, load current 0.11A, capacity 0.004Ah, discharge time 00:02;
Note: If the maximum discharge time is turned on, 00:02 indicates a countdown;
Alarm Status
Upload Alarm status code (OVP/OCP/OPP/LVP/OAH/OHP/OTP)
Read product parameter settings
OVP:25.2, OCP:5.10, OPP:35.5, LVP:1.50,OAH:0222,OHP:00:00,
Overvoltage, overcurrent, over power, undervoltage, maximum capacity, maximum discharge time